Automated ABR Guidelines
NEONATAL HEARING SCREENING AND ASSESSMENT
AUTOMATED AUDITORY BRAINSTEM RESPONSE INFORMATION AND GUIDELINES
FOR SCREENING HEARING IN BABIES (article mirrored from www.nhsp.info/workbook.shtml)
Editor:Steve Mason 1
Contributors:
Clive Elliott 2, Guy Lightfoot 3 , Dave Parker 4, David Stapells5,
John Stevens6 ,Graham Sutton 7, Mike Vidler 8
1. Medical Physics Department , University Hospital, Nottingham,
UK.
2. Audiology Department, Freeman Hospital, Newcastle, UK.
3. Clinical Engineering Dept., Royal Liverpool University Hospital,
Liverpool, UK.
4. Centre for Human Communication and Deafness, University of Manchester,
Manchester, UK.
5. School of Audiology and Speech Sciences, University of British
Columbia, Vancouver, Canada
6. Department of Medical Physics and Clinical Engineering, Central
Sheffield University Hospitals, Sheffield, UK
7. Audiology Department, Royal Berkshire Hospital, Reading , UK.
8. Audiology Department, Birmingham Children's Hospital, Birmingham,
UK.
INTRODUCTION:
Following the publication of the systematic review on Universal
Neonatal Screening1 a national group was formed to discuss some
of the issues of implementation of Universal Neonatal Hearing Screening.
Several meetings of this group have been held to discuss technical
issues and it was felt appropriate to produce consensus documents
on test methodology where this was possible. The click-evoked ABR
(air conduction) was the first of these documents to be published2
and more recently recommendations on the tonepip ABR and bone conduction
ABR have been produced. Since the ABR is widely employed in the
initial screening process, the group has also decided to put forward
information and guidelines regarding the automated collection and
analysis of the ABR as a screening test (automated ABR:
AABR).
SCOPE:
The document sets out to provide information and guidelines for
testing babies in the first few months of life by AABR using primarily
air conduction click stimuli for the purposes of screening for hearing
loss. Clicks are widely employed for neonatal screening and the
majority of reported studies have employed the click stimulus. However
in the future it is possible that other electrophysiological methods,
such as the tone pip ABR, might become an alternative technique.
Equipment that carries out AABR testing is available from some manufacturers.
The test protocols for this type of equipment are usually fixed
and are specific to different manufacturers. Data collection parameters,
implementation of the test, and scoring algorithms will be discussed.
General aspects of ABR methodology have been reported previously2
and will not be included in detail in this document.
DEFINITION OF THE AUTOMATED ABR (AABR):
Recordings of the ABR performed with an highly automated and standardised
procedure for data collection for the purpose of screening for hearing
loss The presence of a response (pass) or absence (refer) at the
screening intensity level of the stimulus is determined primarily
by a clinically proven machine scoring algorithm operating on-line.
IDEAL FEATURES OFTHE AABR SCREENING TEST:
- Easy application and checking of recording electrodes.
- Quick and user-friendly test procedure.
- Portability for flexible implementation.
- Objective (machine) pass and refer test results on each
ear.
- High sensitivity and high specificity.
- Print-out of test results.
- Availability of recorded waveforms for skilled review
and audit.
PATIENT PREPARATION:
Many aspects of the preparation of the baby for testing are similar
to those described in the air conduction click-evoked ABR protocol
(2) and reference should be made to the relevant
sections in that document. Supplementary information is given here.
Test Environment: If AABR testing is performed
outside the designated clinic area, for example on the ward or in
the community, levels of acoustical and electrical interference
must be sufficiently low so as not to influence the results of the
test. Careful selection of the local test area or room may be necessary
in order to achieve satisfactory environmental conditions.
Choice of electrodes and application: Electrodes should
ideally be low cost and disposable in order to meet the demands
of screening large numbers of babies. If re-usable electrodes are
employed then appropriate precautions must be taken to avoid the
risk of cross infection. Since testing is often performed on very
young babies extreme care must be adopted regarding preparation
of the skin for placement of the electrodes. The use of harsh skin
preparation materials should be avoided.
Electrode contact impedance: Acceptable levels
of contact impedance for the surface recording electrodes vary according
to the AABR system being used. Some systems have automated impedance
testing facilities which will not allow the test to proceed unless
values are below specified levels (for example 10k)
Electrode location: In many AABR systems the location
of recording electrodes is similar to that employed in conventional
ABR testing (2) .
STIMULUS
Stimulus type: Typically a click stimulus generated
by an electrical pulse of 100us pulse duration with alternating
polarity. The stimulus is sometimes interleaved between the right
and left ears in order to enable pseudo-simultaneous testing of
both ears.
Stimulus rate: Relatively high stimulus rates are
employed to minimise test time (typically faster than 30 clicks
per second).
Stimulus level: The screening level of the click
stimulus is typically in the range 35dBnHL to 50dBnHL with respect
to normally hearing young adults Any comparison in the performance
of different AABR systems must take into consideration the baseline
calibration of the click stimulus (ppeSPL). The value of ppeSPL
on which the dBnHL is based may be different for different AABR
systems. The recommended value in the air conduction click protocol
is 33dBpeSPL (2).
Earphone: Many types are in use including Telephonics
TDH39/49, insert earphones e.g. type EAR- 3A, and custom designed
ear shells. Before positioning the earphone, the external ear canal
should be checked for any easily removable debris or blockage before
placement of the earphone. Earphones should be carefully positioned
so that the ear canal is not occluded by any excess pressure.
DATA COLLECTION:
The following test parameters for data collection are fixed as
part of the automated test protocol. They must not be changed as
this may invalidate the machine scoring algorithm.
- Gain or sensitivity of the amplifier.
- Level of amplitude artefact rejection.
- Filter bandwidth.
- Acquisition window.
- Number of averaging sweeps.
WAVEFORM ANALYSIS:
The presence or absence of a response in the recorded waveform
is determined objectively using a machine based scoring algorithm.
Statistical and mathematical techniques are typically employed such
as correlation and response to noise amplitude ratio(3),
template methods (4), and Fsp based analysis (5).
The algorithm must be clinically proven in terms of its performance
(eg. sensitivity and specificity) 6,7. The false
negative rate of an AABR screening test can be investigated using
no-sound trials in a relatively small population of babies. The
screening test on each ear is designated a pass or refer depending
on whether or not a response is present. Occasionally, a re-test
decision may be recommended if the result is marginal or if the
test conditions are unreliable. The algorithm should provide a separate
result for each ear so that referral of babies for further testing
can be initiated using either unilateral or bilateral referral criteria.
PRESENTATION OF RESULTS:
The results of the screening test determined by the algorithm
should be clearly presented as a pass, refer or re-test on each
ear separately. Ideally a print-out from the equipment should be
available that can be included in the case notes of the child. On
some equipment the recorded ABR waveforms are displayed and available
as a print-out. This enables a review and audit of the results of
the screening test by a skilled observer. Under no circumstances
should this interpretation be performed by inexperienced personnel
as part of the screening test.
REFERENCES:
1/ Davis A, Bamford J M, Wilson I , et al. A
critical review of the role of neonatal screening in the detection
of congenital hearing impairment. Health Technology Assessment 1997
(1) (10).
2/ Stevens JC et al. (1999) Click auditory brainstem
responses testing in babies- a recommended test protocol. British
Society of Audiology News bulletin, December Issue
3/ Mason SM (1984) On-line computer scoring of
the auditory brainstem response for estimation of hearing threshold.
Audiology 23:277-296
4/ Kileny PR (1998) New insights on infant ABR
hearing screening. Scandinavian Audiology Supplement 30: 81-88.
5/ Sininger Y (2000). Power-optimised Variance
Ratio (POVR). Presented at the Association of Research in Otolaryngology,
February 2000.
6/ Mason SM, Davis A, Wood S, Farnsworth A (1998).
Field sensitivity of targeted neonatal hearing screening using the
Nottingham ABR Screener. Ear & Hearing 19: 91- 102.
7/ Herrmann BS, Thornton AR, Joseph JM. (1995)
Automated infant screening using the ABR: development and validation.
American Journal of Audiology 4: 6-14
Document created 26/11/01 Last modified 12/08/02
|